Twister / Twister Junior
Service Manual
Issued: 21.01.02
Page 157
17.2.3 Performance
test:
•
Pull the actuator motor plug to be tested from the socket
board.
•
Apply a test voltage of 24 V direct-current to the plug
connection check windows.
•
Interchange the testing connections to reverse the
moving direction.
Caution
:
Avoid bridging the contacts!
Danger of fire and combustion !
Fault Cause
Remedy
No continuity can be
measured.
Performance test
negative.
Broken cable.
Motor
defective.
Replace actuator motor.
Replace actuator motor.
Performance
-
+
U
prüf
=24 V