Quality
assurance
User Manual - RAB237AEN
5–7
Control identification
1-
Press "
Lot number
" field to select it and enter the control lot number.
2-
Read the contol label by means of barcode reader or type the control lot number using the virtual
keyboard or an external keyboard (if one is connected).
3-
Press "
Name
" field to select it and enter the control name.
4-
Type the control name in the "
Name
" field using the virtual keyboard or an external keyboard (if
one is connected).
5-
Once the control blood is identified, it has to be set: See
Control setting
In the "
Control Modification
" screen:
1-
Press "
Expiry date
" field to select it to set the control blood expiration date.
2-
Type the control blood expiration date in the "
Expiry date
" field using the virtual keyboard or an
external keyboard (if one is connected).
◆
The expiration date must be given as a date with the date format set in the "
Settings
" menu (See
3.4. Date format setting, page 6-5
3-
Press "
InterQC timelimit
" field to select it and set the maximum time allowed between 2 control
runs using the virtual keyboard or an external keyboard (if one is connected). By default is 24 hours.
◆
The time between two control runs is given in hours.
4-
To validate the control creation: See
Control creation validation, page 5-7
Control creation validation
1-
Press "
Close
" key to close the virtual keyboard if it is opened.
2-
Press "
OK
" button in the "
Control Modification
" screen.
◆
The new control is created. It can be modified as long as no control analysis is running on it.
◆
Your own target values need to be entered.
3-
To enter control target values: See
Entering control target values, page 5-7
.
Entering control target values
In the "
Control Modification
" screen.
1-
Press "
Targets
" button (
◆
The "
Target Modification
" screen is displayed.
2-
Press "
Edit
" button.
3-
Press corresponding radio button to select a parameter family.
Diag.5-6 Target button
Diag.5-7 Target modification screen
Summary of Contents for ABX MICROS ES60
Page 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Page 2: ......
Page 6: ...User Manual RAB237AEN...
Page 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Page 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Page 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Page 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Page 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...