ABX Micros
ES60
OT/CT
7–34
User Manual - RAB237AEN
No results in analysis
Possible dirty apertures
Perform a concentrated cleaning (See
Concentrated cleaning, page 7-9
), rerun
sample
Has diluent been delivered
correctly into RBC chamber?
Check for diluent delivering into chamber,
contact your local HORIBA ABX Technical
Support Representative
Suspicious patient
results
Possible dirty aperture
Perform a concentrated cleaning (See
Concentrated cleaning, page 7-9
), rerun
sample
Possible leakage from syringe
blocks
Check syringe blocks for leakage and contact
your local HORIBA ABX Technical Support
Representative
Possible dirty chamber
Perform a concentrated cleaning (See
Concentrated cleaning, page 7-9
), rerun
sample
QC outside acceptable
limits
Possible dirty aperture
Perform a concentrated cleaning (See
Concentrated cleaning, page 7-9
), rerun
sample
Possible leakage from syringe
blocks
Check syringe blocks for leakage and contact
your local HORIBA ABX Technical Support
Representative
Poor calibration
Recalibrate and rerun QC (See Section 5,
Poor repeatability
Possible dirty apertures
Perform a concentrated cleaning (See
Concentrated cleaning, page 7-9
), rerun
sample
Possible leakage from syringe
blocks
Check syringe blocks for leakage and contact
your local HORIBA ABX Technical Support
Representative
Possible partial blockage in
sampling needle
Remove and flush out sampling needle (See
3.7. Unblocking sampling needle, page 7-17
)
Possible dirty chamber
Perform a concentrated cleaning (See
Concentrated cleaning, page 7-9
), rerun
sample
Incorrect sample mixing
Mix sample just before placing into the tube
holder for each sample (CT). Mix sample just
before pressing the manual sample bar for
each sample (OT).
Problem noted
Possible cause
Corrective action
Summary of Contents for ABX MICROS ES60
Page 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Page 2: ......
Page 6: ...User Manual RAB237AEN...
Page 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Page 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Page 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Page 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Page 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...