ABX Micros
ES60
OT/CT
7–42
User Manual - RAB237AEN
6.2.2. Errors logs
Software, connection and communication
Cannot print result
During the automatic result printout: impossible to convert in
current units or error during report creation or error during
printout:
- Check if printer is on line. If not, switch it on.
- Check if there is enough paper. If not, See
ticket printer paper changing, page 7-16
.
- Check printer settings (See Section 6,
Cannot upload result
LIS sending error:
- Check if LIS connection cable is correctly plugged.
- Restart the instrument.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Invalid QC "lot number"
A QC analysis is out of normality limits:
- Check the last control results.
- Run an analysis for this control (See Section 4,
control blood (ABX Minotrol 16), page 4-11
- Replace the control lot (See Section 5,
XB drift
A statistic dot is out of limits:
- Check XB data.
- Replace the control lot (See Section 5,
Barcode reader error
- Check barcode reader settings (See Section 6,
- Restart the instrument.
Ticket printer error
- Check if ticket printer is online. If not, switch it on.
- Check if there is enough paper.
- Check printer setting (See Section 6,
- Restart the instrument.
Message
Possible cause/Corrective action
Software internal error, try to restart
the application.
Reception of an analysis result instead of a QC result or
conversely.
Internal read/write error.
- Restart the instrument.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Internal read command failure, try to
restart the application.
Internal communication error:
- Restart the instrument.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Internal write command failure, try to
restart the application.
Internal communication error:
- Restart the instrument.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Read write / error on history file, the
application must be reinstalled,
contact your local technical support.
Error during "result historic", or "QC list", or "LJ graphic", or
"LJ save", or «result save» or «result in historic save»
recuperation.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Message
Possible cause/Corrective action
Summary of Contents for ABX MICROS ES60
Page 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Page 2: ......
Page 6: ...User Manual RAB237AEN...
Page 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Page 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Page 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Page 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Page 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...