Maintenance &
Troubleshooting
User Manual - RAB237AEN
7–39
6.1.4. Archive alarms
Low reagent level
DILUENT
Check DILUENT level in Maintenance menu /
Reagent screen. If necessary, replace the
appropriate reagent, run priming cycle from
reagent screen and run a Startup cycle.
DILUENT reagent is almost empty
(counter of tests lower than 10):
- Replace the DILUENT reagent.
See
4.2. Reagent management, page 7-21
Low reagent level
CLEANER
Check CLEANER level in Maintenance menu /
Reagent screen. If necessary, replace the
appropriate reagent, run priming cycle from
reagent screen and run a Startup cycle.
CLEANER reagent is almost empty
(counter of tests lower than 10):
- Replace the CLEANER reagent.
See
4.2. Reagent management, page 7-21
Low reagent level
PACK
Check PACK level in Maintenance menu / Reagent
screen. If necessary, replace the appropriate
reagent, run priming cycle from reagent screen
and run a Startup cycle.
Reagent pack is almost empty (counter of
tests lower than 10):
- Replace the reagent pack.
See
4.2. Reagent management, page 7-21
Reagent empty
Check level in Maintenance menu / Reagent
screen. If necessary, replace the appropriate
reagent, run priming cycle from reagent screen
and run a Startup cycle.
- Replace the pack or the empty reagent.
See
4.2. Reagent management, page 7-21
Diag.7-45 Archive status button
System alarm
message
Help message
Possible cause / Corrective action
settings archive
should be archived.
Archive the settings category in the Maintenance
menu / Archives screen by connecting a USB key
on the analyser.
- Create archive for the software settings.
See
4.3. Archives management, page 7-25
analyser archive
should be archived.
Archive the analyser category in the Maintenance
menu / Archives screen by connecting a USB key
on the analyser.
- Create archive for the Analyser settings.
See
4.3. Archives management, page 7-25
logs archive should
be archived.
Archive the logs category in the Maintenance
menu / Archives screen by connecting a USB key
on the analyser.
- Create archive for the Logs.
See
4.3. Archives management, page 7-25
qc archive should be
archived.
Archive the qc category in the Maintenance menu
/ Archives screen by connecting a USB key on the
analyser.
- Create archive for the QC.
See
4.3. Archives management, page 7-25
analysis archive
should be archived.
Archive the analysis category in the Maintenance
menu / Archives screen by connecting a USB key
on the analyser.
- Create archive for the Patient results.
See
4.3. Archives management, page 7-25
System alarm
message
Help message
Possible cause / Corrective action
Summary of Contents for ABX MICROS ES60
Page 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Page 2: ......
Page 6: ...User Manual RAB237AEN...
Page 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Page 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Page 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Page 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Page 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...