Maintenance &
Troubleshooting
User Manual - RAB237AEN
7–23
Pack unit
In the "
Reag.
" tab:
1-
Press "
View
" button (
).
The "
Edit reagent
" screen is displayed.
2-
Press "
Edit
" button.
3-
Press "
Lot
" field to select it.
4-
Read the reagent pack label by means of barcode reader or type the reagent pack lot number using
the virtual keyboard or an external keyboard (if one is connected).
5-
Press "
Close
" key to close the virtual keyboard if it is opened.
The instrument searches reagent pack information in the reagent database.
When reagent pack information is found, it is displayed as follows:
◆
The reagent pack name in the "
Name
" field.
◆
The reagent pack expiration date in the "
Expiry date
" field.
◆
The reagent volume in the "
Initial volume
" field.
6-
If necessary, the reagent volume can be modified:
◆
Type the reagent volume value in the "
Initial volume
" field using the virtual keyboard or an
external keyboard (if one is connected).
◆
Press "
Close
" key to close the virtual keyboard if it is opened.
7-
Press "
OK
" button to validate.
8-
Press "
Back
" button to return to previous screen.
9-
Install the pack on the instrument (See Section 7,
3.5. Reagent installation, page 7-13
10-
Run a prime cycle (See Section 7,
4.2.4. Reagent priming, page 7-24
).
4.2.3. Reagent replacement
Bottle unit
In the "
Reag.
" tab:
1-
Press corresponding radio button ("
LYSE
", "
DILUENT
" or "
CLEANER
") to select the reagent type
to replace.
2-
Press "
Replace
" button (
).
◆
The following message is displayed: "
Are you sure you want to replace this reagent / pack?
"
3-
Press "
OK
" button.
The "
Edit reagent
" screen is automatically displayed.
4-
Edit a new reagent (See
4.2.2. Reagent creation, page 7-22
).
Pack unit
In the "
Reag.
" tab:
1-
Press "
Replace
" button (
).
◆
The following message is displayed: "
Are you sure you want to replace this reagent / pack?
"
2-
Press "
OK
" button.
The "
Edit reagent
" screen is displayed.
3-
Edit a new reagent pack (See
4.2.2. Reagent creation, page 7-22
).
When a reagent is created, the following information is displayed in the
"
Logs
" tab in the "
Reagent
" category:
- the date/hour of the reagent creation
- the user who has created the reagent
Diag.7-30 Replace button
When a reagent is replaced, the following information is displayed in the
"
Logs
" tab in the "
Reagent
" category:
- the date/hour of the reagent replacement
- the user who has replaced the reagent
Summary of Contents for ABX MICROS ES60
Page 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Page 2: ......
Page 6: ...User Manual RAB237AEN...
Page 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Page 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Page 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Page 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Page 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...