Ticket printer do not answer (timeout)
- Check if ticket printer is online. If not, switch it on.
- Check if there is enough paper.
- Restart the instrument.
Ticket printer initialisation failed.
- Check if ticket printer is online. If not, switch it on.
- Check if there is enough paper.
- Restart the instrument.
Message
Possible cause/Corrective action
Analyser connection failed (internal
RS232).
Internal communication error:
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Analyser com port initialisation failed
Internal communication error:
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Message
Possible cause/Corrective action
Previous invalid QC result automatic
deletion.
Invalid QC result is deleted when directly followed by a QC
passed result.
QC file write error.
Internal read/write error:
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
QC control not found
Impossible to found a QC with this ID.
- Check the control lot number.
Message
Possible cause/Corrective action
Bar code reader initialisation failed
Barcode reader initialization failure in the sample
identification screens.
- Restart the instrument
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Cannot configure bar code reader.
Error during the barcode reader serial port opening
- Restart the instrument
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Cannot connect to bar code reader.
Error during the barcode reader initialization.
- Restart the instrument
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Bar code reader cannot read.
Error during the barcode reader functionning.
Message
Possible cause/Corrective action
Carriage positioning mechanic cycle
failed
Error during carriage motion check cycle
- Manually check the mechanical movement
- Run a startup cycle by pressing the "
Startup
" button.
Diluter positioning mechanic cycle
failed
Error during diluter motion check cycle
- Manually check the mechanical movement
- Run a startup cycle by pressing the "
Startup
" button.
Message
Possible cause/Corrective action
Summary of Contents for ABX MICROS ES60
Page 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Page 2: ......
Page 6: ...User Manual RAB237AEN...
Page 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Page 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Page 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Page 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Page 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...