3.4. Extended concentrated cleaning
Concerns:
◆
RBC/PLT chamber and aperture cleaning
◆
WBC/HGB chamber and aperture cleaning
Required tools:
◆
5ml or 10ml syringe
Required products:
◆
ABX Minoclair 0.5L (P/n: 0401005) or bleach containing 12% active chlorine
Intervention time:
◆
Approximately 30 minutes
Specific kits or consumable products:
◆
None
Procedure:
1-
Perform 3 backflush cycles (See
).
2-
Drain the chambers (See
3.3.4. Chambers draining, page 7-9
3-
Open the main door (See
3.2. Cover dismantling, page 7-6
).
4-
Locate the opening on the top of the chamber (shield or not).
5-
Fill both chambers with 5mL of the ABX Minoclair.
6-
Let the solution set for 10 minutes in the chambers.
7-
After 10 minutes, run a concentrated cleaning (See
3.3.5. Concentrated cleaning, page 7-9
).
8-
Once the cleaning cycle is complete, perform 3 backflush cycles (See
9-
Press "
Status
" button to open "
Status
" area.
10-
Press "
Startup
" button to run a startup. Verify that reference blank results are within their limits.
11-
Re-run quality control and verify that the results are within the assay limits listed on the control
assay sheet that comes with the controls.
◆
If the problem still persists, repeat steps <1> through <11> again.
◆
If the problem still persists after the second cleaning, call your local HORIBA ABX Technical
Support Representative.
Summary of Contents for ABX MICROS ES60
Page 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Page 2: ......
Page 6: ...User Manual RAB237AEN...
Page 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Page 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Page 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Page 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Page 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...