ABX Micros
ES60
OT/CT
7–44
User Manual - RAB237AEN
Send result failed.
- Check if LIS connection cable is correctly plugged.
- Restart the instrument.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Minor error (key not found in %s).
Internal read/write error.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Invalid patient type, software internal
error.
Internal read/write error.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Unexpected error (%ld)
Internal read/write error.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Settings file not found
Internal read/write error.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Settings file corrupted
Setting file is corrupted.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Settings file not found : Translations
Setting file is missing.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Incompatible EPROM version.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Archive restoration failed
Internal read/write error.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Online help installation failed, check
the help file on your USB key.
USB key removal during reading.
Missing or corrupted files on USB key:
- Check the help files on your USB key.
LIS connection failed
- Check if LIS connection cable is correctly plugged.
- Restart the instrument.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Error while deleting user.
Internal read/write error.
- Restart the instrument.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Backup of current default reagents
configuration files failed
Internal read/write error.
- Restart the instrument.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Invalid reagents import file format
Internal read/write error.
- Restart the instrument.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Copy of default reagents files failed
Internal read/write error.
- Restart the instrument.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Invalid default reagents import file
Internal read/write error.
- Restart the instrument.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Message
Possible cause/Corrective action
Summary of Contents for ABX MICROS ES60
Page 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Page 2: ......
Page 6: ...User Manual RAB237AEN...
Page 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Page 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Page 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Page 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Page 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...