Maintenance &
Troubleshooting
User Manual - RAB237AEN
7–13
3.5. Reagent installation
3.5.1. Bottle connections
◆
Lyse and cleaning reagents are placed inside the reagent compartment (
1-
Install the reagent straws and the bottle stoppers on the reagent bottles.
2-
Connect the blue tube to the ABX Cleaner bottle and the white tube to the ABX Minilyse LMG
bottle.
3-
Close the compartment cover.
4-
to run the priming of the replaced reagent.
3.5.2. Alphalyse 360
Required products:
◆
Lyse straw (P/n: G0166740). Contact your local HORIBA ABX Technical support to order it.
1-
Open the pack pulling the flap A and open the level window B (
).
2-
Replace the rigid straw by the Lyse straw and install the Alphalyse 360 as shown on
.
◆
Make sure that a blank cycle and control run will be carried out after the
change of a reagent in the course of day.
◆
Before reagent installation, register reagent pack or bottles in the "
Reag.
"
tab (See
4.2.2. Reagent creation, page 7-22
).
Diag.7-16 Reagent bottles
Diag.7-17 Opening
B
A