Configuration
85
•
•
•
•
•
•
On the 4050 display
:
Step 1.
Press the down arrow to select Low Pass Filter.
Step 2.
Press the right arrow to toggle between ON or OFF.
Step 3.
Press Enter to save your choice.
Motion Tolerance Parameter
The value you enter for Motion Tolerance sets the amount of deviation to allow for your
process. This value must be greater than or equal to the Graduation Sizes. We recommend
3 graduation sizes
The base motion number can be calculated by using the following formula:
Base Motion Number = (Total Load Cell Capacity x 0.0003)
Motion Tolerance must be greater than or equal to the Graduation Sizes. We recommend
three graduation sizes.
RANGE: .000001 - 999999 (default 10)
On the Web page
, enter the value in the text box provided.
On the 4050 display:
Step 1.
After selecting Motion Tolerance with the down
arrow on the Setup menu, press Enter to display
the Motion Tolerance edit form.
Step 2.
Use the left or right arrow to position the cursor
and the up or down arrows change the value
Step 3.
When you have selected all the values, press Enter to save.
Step 4.
Use the left or right arrow to position the cursor and the up or down arrows change
the value. Our example uses .05, which is for illustration only. Yours will vary
depending on your application.
Step 5.
When you have selected all the values, press Enter to save.
Zero Tolerance Parameter (Gross Weight)
The value you enter for Zero Tolerance sets the weight unit limit from zero the instrument
will accept as gross zero during the zeroing function (when you push the Zero button).
RANGE: .000001 - 999999 . (default 10.0)
On the Web page
, enter the value in the text box provided.
Our example, for illustration only, sets Zero Tolerance to 10.
On the 4050 display: