Configuration
73
•
•
•
•
•
•
Step 4.
Complete any additional configuration that is required by your PLC for
initialization. Our initialization example is for a Siemens PLC. Your PLC
initialization requirements may differ.
Step 5.
Install the *.GSD file for the instrument you connected to the Profibus Network.
Initialization Procedures
NOTE
The examples come from the Siemens Step 7™, Simatic Manager Software. Your software
will vary from these procedures. Step 7™ is a trademark of the Siemens Corporation.
Step 1.
In the Siemens Step 7™, Simatic
Manager open the Hardware Catalog.
Step 2.
Click on the “+” to expand the
Additional Field Devices Folder.
Step 3.
Click + to expand the General Folder.
Step 4.
Highlight the CPU you selected in the
UR dialog box.
Step 5.
Double Click on “HI4000” or drag and
drop the “HI4000” folder to the
Profibus-DP Network.
Step 6.
This opens the HI 4000 PDP
Parameters dialog box where you can
set the address of the instrument, if
necessary.