78
Chapter
4
•
•
•
•
•
•
On the Web page
, enter a name to identify the
instrument in the Instrument ID text field. We used
“HI 4050.”
On the 4050 display
, use the
Down arrow to select Instru-
ment ID, then press Enter to
display the Instrument ID edit
page.
Use the left or right arrow to position the cursor and the up or down arrows change the value
Enter saves the entry.
Operator ID Parameter
The Operator ID is the ID of the user who is going to operate the Weight Controller or
service the instrument. Select three letters or numbers or any combination of letters and
numbers that adequately identifies the user.
On the Web page
, type a name that identifies the
operator of this instrument in Operator ID text field.
On the 4050 display
:
Step 1.
Press the Down
arrow to select
Operator ID.
Step 2.
Press Enter to open
the Operator ID edit
screen.
Step 3.
Use the left or right arrow to position the cursor and the up or down arrows change
the value
Step 4.
When you have selected all the values, press Enter to save.
Decimal Point Parameter
Step 5.
Use the Decimal Point Parameter to set the resolution you want for the instrument.
Here you set the location of the decimal point for the weight resolution. The
higher the number, the farther to the left the decimal point moves and the higher
the resolution of the scale. Note that setting more decimal points does affect the
overall accuracy of the instrument. Adding decimal points will not improve a
system accuracy beyond the specified ability of the load cell.
RANGE: 0-5 (default 2)
>
_
>
_