106
Chapter
5
•
•
•
•
•
•
Step 4.
Click the Do Cal Low button to do the
Trad Cal Zero. A “Cal Completed OK”
message appears briefly if the calibra-
tion was successful. An Error number
appears if the calibration was not
successful. See the Error list in Chapter
8 for help in correcting the error.
Step 5.
To set the Span Weight, place a certified
test weight on the scale.
Step 6.
In the Span Weight text field, enter the
amount of live load weight placed on
the scale.
Step 7.
Click the Do Cal High button. A “Cal
Completed OK” message appears briefly if the calibration was successful.
Step 8.
An Error message appears if
the calibration was not
successful. See the error list in
Chapter 8, for help in
correcting errors.
Step 9.
The scale is now calibrated.
Traditional Calibration from the Front Panel
Step 1.
From the Configuration menu, use the down arrow to select Calibration, then
press Enter to display the Calibration menu. Select Trad Cal and press Enter to
open the Trad Cal menu; then press Enter to open the Reference Weight edit form.
Step 2.
Set the reference weight equal to the weight on the scale. Normally, you would
remove all weight “live load” from the Scale to obtain a Ref Weight of 0.0. Wait
12 seconds or more for the reading to stabilize.