Calibration
103
•
•
•
•
•
•
C2 Calibration From the Web Page
From the Home Page, click Configuration
to display the Configuration menu; then
click Calibration to open the Calibration
page.
Step 1.
In the “Ref Weight” text field,
enter the reference weight for
your application.
Step 2.
(optional) If you want
to enter the Gravity
Correction Factor, see
the table above, then
enter the Correction
Factor number you
selected from the table
in the Gravity
Correction field .
Step 3.
Click the “Do C2
Calibration” button.
Step 4.
Wait a few seconds and the results
will appear. If the Calibration was
successful a Cal completed OK
appears and the Load Sensor
Number will read 0.
Step 5.
This message appears if no,
incorrectly wired, or over ranged
load points are connected to the
4050. Check the cable and
connectors of the load point(s) and re-calibrate.