30
FMS 1 - Fluorescence Monitoring System
© 2017 Hansatech Instruments Ltd
Fig. 2-4
. During the saturation pulse (marked by the arrows) the fluorescence signal
exceeds the range of the A/D converter. The peak of the pulse is cut from the trace
but sampled (incorrectly) as 4095 bits by the software.
The peak fluorescence signal is effectively cut from the recording. This situation may be corrected
by reducing instrument gain (
) or re-orienting the fibre-optic further away from the
sample. If left un-corrected erroneous measurements of Fm and PSII will be stored when
measurements are made.
As referenced above, all of the FMS configuration and data management facilities are accessed
from the “drop-down” menu (C, Figure 2-1) whilst execution of measurements and on-line control
of instrument light sources is initiated via the
instrument toolbar (D, Figure 2-1). The following sections detail the configuration and data
management facilities available from the “drop-down” menu. Description of each item on the
instrument toolbar is given in
3.3
Drop-Down” Menu Functions
The “drop-down” menu contains seven groups of related functions under the
following headings:
Files Script Hardware Graph Tools View Help
In this section the groups are described in three broad catagories:
Summary of Contents for FMS1
Page 1: ......
Page 7: ...Section 1 Measurement Principles and Hardware ...
Page 21: ...Section 2 Software Installation System Operation Data Handling ...
Page 70: ...70 FMS 1 Fluorescence Monitoring System 2017 Hansatech Instruments Ltd ...
Page 71: ...Section 3 Stand Alone Instrument Operation and Data Upload ...
Page 83: ...83 Section 3 Stand Alone Instrument Operation and Data Upload 2017 Hansatech Instruments Ltd ...