Number English Operator Message
Translated Operator Message
33
There are still scan tasks to be completed. Done Scanning
will end the scanning portion of this exam. Are you sure you
are done scanning?
34
Tube temperature is too cold. Warm-up is needed.
35
Attention
An unrecognized tube has been installed on the system.
Fast Calibration techniques are designed specifically for
GE tubes and GE cannot guarantee that the performance
of the system will meet specifications with an unrecognized
tube.
36
An unrecognized X-Ray tube has been installed on the sys‐
tem.
GE cannot assure that system performance will conform to
specifications.
Advisory messages will be posted to the operator about an
unrecognized tube during tube warm-up, during Fast Cali‐
bration, and in the dose report.
37
Data transfer error.
Please refresh patient schedule and retry. If problem per‐
sists contact your service representative.
38
Cannot start a New Patient scan session while in an active
scan session.
Close if any of the following application(s) are running:
39
Cannot switch patient sessions while on the scan progress
screen.
Close if any of the following application(s) are running:
40
This scan prescription has selected one of the options for
Recon based ACR Image Flip and Rotate.
Please ensure that this desired orientation is displayed ap‐
propriately on all remote viewing devices.
41
Maximum number of exam sessions are currently open.
Please complete, as needed, at least one exam session.
Once a session is closed, start a new patient session.
Table 4: Table Operator Messages
Number English Operator Message
Translated Operator Message
1
Table received an invalid request. Please retry positioning
the table.
If problem persists, select Done Scanning and perform a
Scan Hardware Reset from the Service Tools menu.
2
Table did not reach position. Please retry positioning the ta‐
ble.
If problem persists, select Done Scanning and perform a
Scan Hardware Reset from the Service Tools menu.
3
Table motion is unavailable.
Please perform a Scan Hardware Reset from the Service
Tools menu.
If problem persists, contact your service representative.
4
Table motion is unavailable.
Please select Done Scanning and perform a Scan Hard‐
ware Reset from the Service Tools menu.
If problem persists, contact your service representative.