4 Configure the System for Dose Check
Use this procedure to configure the system for dose checking by Notification Value (NV) and
Alert Value (AV). You must have one user assigned to a group with the Dose Administrator role
to confirm dose check settings.
1. From the image monitor, click .
2. From the
Mode
menu, click
Access Controls
, then
Dose Check Management
.
3. From the
Dose Check Management
window, in the
NV (Notification Value) Checking
area,
select
CTDI
vol
and/or
DLP
.
NOTE:
The system default for Notification Value (NV) Checking is
CTDI
vol
On
and
DLP
Off
. The GE Reference protocols will not contain any NV values.
Illustration 1: Dose Check Enabled
4. From the
Dose Check Management
window, in the
AV (Alert Value) Checking
area, select
CTDI
vol
and/or
DLP
.
NOTE:
The Alert Value (AV) should be determined by the site’s radiologist and physicist.
5. Enter a value for
CTDI
vol
and/or
DLP
.
○
The maximum AV value for CTDI
vol
is 2,000 mGy.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Chapter 7 Dose Check