Table 1: mA adjustment factor
ASiR-V
Normal
10%
0.87
20%
0.74
30%
0.63
40%
0.52
50%
0.44
60%
0.35
70%
0.29
80%
0.23
90%
0.18
100%
0.14
For example, if 40% ASiR-V is selected, multiply the mA displayed in the mA control screen by
to obtain the new mA for protocol using ASiR-V, that needs to be
entered into the mA field on the mA control screen.
Multiply the Noise Index or manual mA values that were adjusted for an ASiR-V applied protocol
.
Use the following factors to return any manual mA values adjusted for ASiR-V to non- ASiR-V
value.
Table 2: Return to a value for a non-ASiR protocol
ASiR Level
Return to non-ASiR manual mA value
30%
1.59
40%
1.92
50%
2.29
2.3 Reconstruction
Reconstruction of ASiR-V images involves defining the level of noise reduction desired for the
parameters used for various applications. Image reconstruction is a blending of the original
images and a percentage of an ASiR-V image. There are 10 blending levels available that are
the amount of noise reduction based on the maximum of the 100% image reconstructed with the
original image data. Reconstruction of ASiR-V images can be prescribed retrospectively in
Retro Recon. A New Series should be selected for each reconstruction of the data with ASiR-V
applied so comparisons and identification of series with ASiR-V can be made more easily.
ASiR-V is located on the Primary and secondary recon settings collection. ASiR-V Setup
Selection is 0% to 100%.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
310