6.2.12
Specimen Slide Removal / Replacement
INTRODUCTION
The specimen slide controls the vertical position of
the specimen relative to the x-ray source.
There are 2 methods of removing the specimen
slide:
A.) Remove rotary stage first, then only remove the
specimen slide, OR
B.) Remove the entire rotary stage and specimen
slide assembly (together), and then separate
the specimen slide from the rotary stage (this
method may be easier than separating the
specimen slide from the rotary stage in the
small confines of the open door).
Both methods are described below.
TOOL LIST
Standard toolkit
SUPPLIES
Cable ties
SAFETY CONSIDERATIONS
X-ray shielding contains lead and lead compounds, chemicals known (to the State of California) to
cause cancer and other reproductive harm. Wear gloves when handling shielding and wash hands
after handling.
PRECONDITIONS
•
Machine locked out. See 6.1.1 for details.
•
Left access panel removed. See 6.1.2 for details.
•
Equalization bath bracket removed. See 6.2.7 for details.
•
Rotary stage removed OR entire rotary stage and specimen slide assembly.
Specimen slide
assembly
Summary of Contents for Healthcare eXplore Locus SP
Page 2: ...This page is intentionally left blank...
Page 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Page 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Page 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Page 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...