Capture an image and name it
D1
.
5.
Move the tool back 100mm
TO THE RIGHT
(toward the X-Ray Source).
6.
Capture an image and name it
D2
.
7.
Turn off the X-Ray Source and remove Distance tool from the scanner.
8.
Turn the X-Ray Source back on to keep it warm.
9.
Correct and unwarp the D1 image and load it into
MicroView
.
10. Draw an ROI (Region Of Interest) in
MicroView
in the middle of the white area:
-
Position cursor in the white area, near the centre, and press the
7
key. This draws the top
left corner of the ROI.
-
Move cursor to the right and down and press the
8
key to complete the ROI. A yellow box
represents the ROI.
-
Press small
m
(no caps).
The mean value is displayed in the bottom right corner of the window. Make a note of this
value in the Worksheet.