GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 87
3.
Using 1.5mm hex key, remove 3 screws
securing shutter motor to bracket.
REPLACE
1.
Reversal of removal.
2.
Ensure that set screw is seated on flat of shaft.
3.
Ensure that shutter blades are centered in
shutter control assembly upon installation.
FOLLOW-ON PROCEDURES
Replace all shielding, covers, case and panels, and
perform an "abridged" calibration (see Chapter 7).
Center shutter
blade in detector