Page 57
6.1.3
Right Access Panel Removal / Replacement
INTRODUCTION
The right access panel needs to be removed for
many of the procedures in this section. Instructions
follow.
TOOL LIST
Standard toolkit
SUPPLIES
None
SAFETY CONSIDERATIONS
X-ray shielding contains lead and lead compounds,
chemicals known (to the State of California) to cause
cancer and other reproductive harm. Wear gloves when handling shielding and wash hands after
handling.
PRECONDITIONS
Machine locked out. See 6.1.1 for details.
REMOVE
1.
Using Phillips screwdriver remove twelve
screws and washers from right panel.
2.
Disconnect ground cable to panel.
3.
Remove panel from top. Lift up panel to right to
remove.
REPLACE
Reverse of removal.
FOLLOW-ON PROCEDURES
None
Right
panel
Front view