Using 5/64” hex key, remove the ground cable
from detector.
5.
Using 9/64” hex key, remove four screws and
lockwashers securing detector to stand.
Support detector while removing. Set aside.
REPLACE
Reverse of removal.
FOLLOW-ON PROCEDURES
Detector Replacement calibration is required (see Chapter 7).