GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 210
18. Adjust the tip on the Precision Dial Indicator to strike the equator of the ball. Make sure the
Indicator is not pegged (i.e. the small needle on the gauge, near the centre of the Dial Indicator,
is around the halfway mark).
Note:
The Precision Dial Indicator is very sensitive and care should be taken not to lean or touch any
part of the system while adjusting the Indicator or the X,Y Centring Ball tool.
19. Zero the Precision Dial Indicator.
Note:
For fine measurements, the Precision Dial Indicator may not be accurate until after the first
movement.
20. Use the
Rotation
setting in
X-Ray and Motor Controller
to rotate the ball 360° (180° at a time).
Watch as the ball is rotating to ensure that the tip of the Precision Dial Indicator is touching the
ball at all points of rotation and that it is not pegged. Re-zero the Dial Indicator, if necessary.
Centre the Ball:
21. Use the
Rotation
setting to rotate the ball 180°.
22. Watch the Precision Dial Indicator as the ball rotates. Note the reading of the Precision Dial
Indicator after 180° of rotation (e.g. .0004").
The objective is to adjust the ball so that, after the adjustment, the Dial Indicator reads half of
the original reading (e.g. .0002").
To do this:
23. Use the tool’s adjustment key to carefully turn the 2 adjustment screws that are aligned with
the Dial Indicator - turn the screws until the Dial Indicator reads half of the original reading.
NOTICE
Equipment
Damage
Possible
The tool is extremely delicate. Fine
adjustments are required. DO NOT
overtighten the screws. One screw
turns clockwise and the other turns
counter clockwise. Turn the screws in
tandem using as little torque as
possible (i.e. 5° at a time).
24. Once adjusted, rotate the ball 180°
and check the Dial Indicator reading.
It should match the reading taken
after the adjustment in the previous
step (e.g. .0002").
25. Re-zero the Dial Indicator.
Summary of Contents for Healthcare eXplore Locus SP
Page 2: ...This page is intentionally left blank...
Page 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Page 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Page 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Page 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...