Page 107
4.
Close
MicroView
upon
completion.
5.
Use
Windows Explorer
to
rename the corrected image to
corrected_grid.vff
6.
Run the
transgen
program. The
transgen
program is located in
C:\Program Files\GE Medical
Systems\eXplore
Utilities\Calibration\.
7.
Load the
corrected_grid
file into
the
transgen
program.
8.
Adjust the
Window
(e.g. 2300)
and
Level
(e.g. remove graininess) until there are
minimal pixels visible except for the grid’s bebes.
Make the bebes as big and round as possible,
zooming in on individual bebes to check their
roundness and the surrounding area for visible
stray pixels.
Not getting the Window and Level right may affect
the centroid calculations.
To Pan, move the mouse while pressing Shift plus
the left mouse button simultaneously. To Zoom,
move the mouse up (to magnify) or down (to zoom
out) while pressing the right mouse button.
Note:
If stray pixels touch or are close to any bebe, take a new image.
9.
Place the cursor over the centre of the first bebe on
the top left of the grid and press
1
.
10. Place the cursor over the centre of the last bebe on the top right of the grid and press
2
.
11. Determine how many pixels there are between 1 and 2 (marker 2’s x position minus marker
1’s x position) and record in the Worksheet.
1 . . . . . . . . . 2
. . . . . . . . . . .
. . . . . . . . . . .
. . . . . . . . . . .
. . . . . . . . . . .
Summary of Contents for Healthcare eXplore Locus SP
Page 2: ...This page is intentionally left blank...
Page 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Page 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Page 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Page 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...