26. Select a region of interest that encompasses all 4 wires using the 7 and 8 keys in
MicroView
.
27. Select File>Save crop coordinates.
28. Exit
MicroView
.
29. Click the
Load Crop
button on the
Reconstruction Utility
window.
30. Press the
Recon
button. View the image in
MicroView
and verify the image is cropped
properly, then exit
MicroView
.
31. Click the
Full Res b
utton. De-select the Mini-volume field. This allows you to name the resulting
reconstructed volume.
32. Change the Z-Size field to
1
.
33. Press the
Recon
button. Name the file
1x1wire_cs_ss.vff
(or
2x2wire_cs_ss.vff
or
4x4wire_cs_ss.vff
depending on which verification scans are being performed).