GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 148
7.
Ensure the fields below are set to the following settings:
Enable Left Limit: checked off
Enable Right Limit: checked off
Camera:
Thomson 8M
Bin Mode:
1 x 1
Exposure Time:
3000 ms
kV:
80
µA:
80
8.
Find the motor limit switch (this ensures that the limit switches will never be triggered):
-
select
Magnification
from the drop down menu in the Motor field
-
press the
Left (CW) Linear Home
button to move the stage toward the detector
-
enter
2
in the mm field, click on the
right or CCW
radio button, then press the
Linear Stage
Motion
button to offset the "home" position by 2 mm.
9.
Click the
Start Shutter
button.
10. Click the
Turn X-Ray
On button.
11. Click the
Fluoro
button, then:
-
select
OK
when the warning pop-up appears.
-
select
OK
when the text concerning placement of the specimen appears.