6.
Loosen the 4 detector mounting bolts to adjust the detector alignment first.
7.
Shim the detector mounting bracket to adjust the elevation and twist the detector while the
bolts are loose to adjust the azimuth.
8.
If the alignment can not be accomplished
with this adjustment, loosen and adjust using
the 4 bottom detector mounting bolts.