ecoFLOW
This feature provides a split screen view that shows the approximate
minimum O2 flow to maintain a preset inspired O2 concentration.
Also shown is the approximate agent used per hour and the cost.
1.30
1
2
9
15
-20
100
0
20
40
60
80
AC.22.041
3
100
10
4
0.60
21.91
Agent
Total Flow
O2 Total
Iso
Fi25 O2 flow
4
2
1
3
9
10
8
7
6
5
1.
Agent
Shows Agent cost and flow information.
2.
Agent cost
The cost of the current agent flow. This value is determined by the agent flow multiplied
by the agent cost set in Super user mode.
3.
Agent flow
The measured value of the liquid agent flow from the vaporizer. The agent flow may have
a delayed response. For example: Iso.
4.
FiO2 flow marker
The graphical representation on the flow tube of the FiO2 flow value. This marker can be
removed by disabling it in Super user mode.
5.
FiO2 flow
The minimum O2 flow needed to maintain the set inspired O2 flow. This item can be
disabled in Super user mode. For example: Fi25 O2 flow.
6.
O2 total (numeric)
The numeric representation of the total O2 flow. If N2O is the balance gas, this equals the
set O2 flow. If Air is the balance gas, this is the set O2 flow plus 21% of the Air flow. For
example: O2 Total.
7.
O2 total (graphical)
The graphical representation of the total O2 flow. If N2O is the balance gas, this equals
the set O2 flow. If Air is the balance gas, this is the set O2 flow plus 21% of the Air flow.
8.
Remaining gas flow
If N2O is the balance gas, this equals the set N2O flow. If Air is the balance gas, this N2
is 79% of the Air flow.
9.
Flow bobbin
The height of this represents the total fresh gas flow delivered to the patient.
10. Total flow
Shows Total Flow information.
Figure 3-3 • ecoFLOW feature
Aisys CS²
3-32
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...