Super user mode
Do not enter Super User mode when a patient is
connected to the system. Gas flow will cease, and the
system must be powered down in order to restart gas
flow.
Changes made in Super User mode affect the system
configuration. All changes made are permanent and
preserved until changed again.
Several settings can be changed in Super User mode. These
changes should only be made by the person responsible for the
configuration of the system.
To access the Super User mode, select
System Setup
and then
Super User
. The Super User mode is password protected. Contact
the sales representative to obtain the password.
The system must be turned off to exit Super User mode.
Menu items are dependant on system configuration. Inactive menu
items are gray and are not selectable.
The menu selections shown in the examples are the factory default
values. The available settings are shown to the right of the menu.
WARNING
•
Note
Note
Aisys CS²
12-2
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...