Message
Priority
Cause
Action
Increase low MV
limit
Medium
Low MV limit is off in SIMV VCV, SIMV
PCV, SIMV PCV-VG, CPAP + PSV, or
PSVPro modes.
Increase low MV alarm limit to improve
patient disconnection detection.
Insert cassette
Low
Cassette removal detected during
active delivery.
Reinsert the cassette.
Inspiration
stopped
Medium
High airway pressure.
Check system for blockages.
Internal failure.
System may shut
down.
High
Power controller software failure.
Contact an authorized service
representative.
Internal failure.
System may shut
down.
Medium
Power controller software failure.
Contact an authorized service
representative.
Low gas sample
flow
Low
Sample flow is less than 80% of
nominal flow for 20 seconds.
Check for blockage in the airway
module sample gas line.
Memory
(EEPROM)
failure
Low
Software error.
Contact an authorized service
representative.
Module fail. No
CO2, AA, O2
data.
Medium
Airway module hardware failure.
Replace module.
Module not
compatible
Low
The monitoring module detected is not
compatible with system software.
Remove the incompatible module. Use
a compatible module.
Move Bag/Vent
switch to Bag
Medium
Bag/Vent switch was in the Vent
position when the case was started.
Move switch to the Bag position.
MVexp high
Medium
MVexp is greater than MVexp high
alarm limit (for nine breaths and one
minute has elapsed).
Set the alarm limits appropriately.
Check the ventilation settings.
MVexp low
Medium
MVexp is less than MVexp low alarm
limit (for nine breaths and one minute
has elapsed).
Set the alarm limits appropriately.
Check the ventilation settings.
N2O supply
pressure low
Medium
N2O pipeline pressure is less than 252
kPa (36 psi) and the N2O cylinder
pressure is less than 2633 kPa (381
psi).
Ensure the N2O pipeline and cylinder
are properly connected.
Negative airway
pressure
High
Paw is less than -10 cmH2O.
Check for blockages in the patient
circuit.
No battery
backup
Medium
Battery or charging failure.
Between cases turn the system circuit
breaker off, then back on after 15
seconds to reset the system.
No exp flow
sensor
Medium
Electrical signals show the flow sensor
is not connected.
Connect the flow sensor. Replace the
flow sensor if necessary.
No fresh gas
flow!
High
Possible patient detected while system
is in checkout state.
Disconnect the patient or start the case.
No fresh gas
flow?
High
Possible fresh gas flow occlusion or
loss of gas pressure.
Switch to circle circuit or bag the
patient. Check pipeline supply
connection.
No insp flow
sensor
Medium
Electrical signals show the flow sensor
is not connected.
Connect the flow sensor. Replace the
flow sensor if necessary.
O2 flush stuck
on?
Low
Switch detected “on” continuously for
more than 30 seconds.
Check flush valve. Ensure flush valve is
not sticking.
7 Alarms and troubleshooting
2067226-001
7-9
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...