3.
Select
Patient
to have spirometry data sourced from the airway
module or
Vent
to have spirometry data sourced from the
ventilator.
4.
Select
Back
to view changes made and access other functions
of the
Spirometry
menu.
Setting sweep speed
Use the
Sweep Speed
setting to set the waveform draw rate to fast
(6.25 mm/s) or slow (0.625 mm/s). When the sweep speed changes,
waveforms redraw at the new rate.
1.
Select
System Setup
-
Screen Setup
.
2.
Select
More Settings
.
3.
Select
Sweep Speed
and then select
Fast
or
Slow
.
4.
Select
Close
.
Setting display brightness
Use the brightness setting to adjust the contrast level of the display.
1.
Select
System Setup
-
Screen Setup
.
2.
Select
More Settings
.
3.
Select
Display Brightness
.
4.
Select the desired brightness level with 1 being the dimmest and
5 being the brightest.
5.
Select
Close
.
Setting keypad brightness
Use the brightness setting to adjust the contrast level of the hard
keys on the bezel.
1.
Select
System Setup
-
Screen Setup
.
2.
Select
More Settings
.
3.
Select
Keypad Brightness
.
4.
Select the desired brightness level with 1 being the dimmest and
5 being the brightest.
Set to 0 to turn off the keypad brightness.
5.
Select
Close
.
3 Operation
2067226-001
3-15
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...