Gas setup
Use the
Gas Setup
menu to adjust the Agent, O2% and total flow, to
change the balance gas, and to change the circuit type.
Changing gas settings
1.
Select the
Gas Setup
quick key.
2.
Select the setting to change from the
Gas Setup
menu.
3.
Change the setting.
4.
For
Circuit
, select the menu item and change using the drop-
down menu. Select
Confirm
.
5.
For
Other Gas
, select the menu item and change using the
drop-down menu.
6.
For
Agent
,
O2%
, and
Total Flow
, select the setting and make
the change using the ComWheel and push to confirm the
setting.
Changing balance gas
1.
Select the
Gas Setup
quick key.
2.
Select the
Other Gas
menu item.
3.
Select the balance gas to use with O2.
Changing circuit type
1.
Select the
Gas Setup
quick key.
2.
Select
Circle
or
Non-Circle
3.
Select
Confirm
.
Aisys CS²
3-10
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...