Message
Priority
Cause
Action
EtSev high
Medium
EtSev is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes.
Set the alarm limits appropriately.
Check agent setting.
EtSev low
Low
EtSev is less than alarm limit.
Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
FiCO2 high.
Absorbent OK?
High
FiCO2 is greater than alarm limit.
Check if absorbent needs to be
changed.
FiDes high
Medium
FiDes is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes.
Set the alarm limits appropriately.
Check agent setting.
FiDes low
Low
FiDes is less than alarm limit.
Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
FiEnf high
Medium
FiEnf is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes
Set the alarm limits appropriately.
Check agent setting.
FiEnf low
Low
FiEnf is less than alarm limit.
Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
FiHal high
Medium
FiHal is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes
Set the alarm limits appropriately.
Check agent setting.
FiHal low
Low
FiHal is less than alarm limit.
Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
FiIso high
Medium
FiIso is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes
Set the alarm limits appropriately.
Check agent setting.
FiIso low
Low
FiIso is less than alarm limit.
Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
FiO2 high
Medium
FiO2 is greater than high alarm limit.
Check the O2 setting. Recalibrate the
O2 cell and the airway module.
FiO2 low
High
FiO2 is less than low alarm limit.
Check O2 setting. Check for leaks or
blockages in the patient circuit.
FiSev high
Medium
FiSev is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes
Set the alarm limits appropriately.
Check agent setting.
FiSev low
Low
FiSev is less than alarm limit.
Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
For mech vent,
set Bag/Vent
switch
Low
Bag/Vent switch is set to Bag and the
ACGO switch is set to Circle.
Move switch to Vent position to start
mechanical ventilation.
Gas monitoring
not available
Medium
Airway module hardware failure.
Replace airway module. Then, turn
power off and back on between cases
to clear the alarm and receive module
data.
Aisys CS²
7-8
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...