Radiated immunity
Immunity test
IEC 60601-1-2 test level Compliance level
Electromagnetic environment
guidance recommended
separation distance
Conducted RF IEC
61000-4-6
3 Vrms
150 kHz to 80 MHz
outside ISM bands
10 Vrms (V1)
D=0.35√P
10 Vrms
150 kHz to 80 MHz in
ISM bands
10 Vrms (V2)
D=1.2√P
Radiated RF IEC
61000-4-3
10 V/m
80 MHz to 2.5 GHz
10 V/m (E1)
D=1.2√P 80 MHz to 800 MHz
D=2.3√P 800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and D is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should
be less than the compliance level in
each frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
•
The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz,
13.553 MHz to 13.567 MHz, 26.957 MHz to 27.283 MHz, and 40.66 MHz to 40.70 MHz.
•
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range
80 MHz to 2.5 GHz are intended to decrease the likelihood that a mobile/portable communications equipment
could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of
10/3 has been incorporated into calculating the recommended separation distance for transmitters in these
frequency ranges.
•
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the system is used
exceeds the applicable RF compliance level above, the system should be observed to verify normal operation.
If abnormal performance is observed, additional measures may not be necessary, such as re-orienting or
relocating the system.
•
Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
11 Specifications and theory of operation
2067226-001
11-49
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...