3.
Set
Leak Audio
to
Off
.
The audio alarms for small leaks are silenced.
4.
To turn audio alarms back on, set
Leak Audio
to
On
.
5.
Select
Close
.
If the Low MV alarm limits are off or
MV/TV Alarms
is set to
Off
,
Leak Audio
is automatically set to
On
and cannot be changed.
--
Setting auto MV limits
MV alarm limits can be calculated automatically for mechanical
ventilation when in VCV or PCV-VG modes and volume
compensation is enabled. Use the
Auto MV Limits
setting to turn on
automatic calculations of the MV alarm limits.
If the automatic calculation of the low or high minute volume alarm
limit exceeds the allowable limit, the minimum or maximum alarm
limit is used.
1.
Select
Alarm Setup
.
2.
Select the
Configure
tab.
3.
Set
Auto MV Limits
to
On
.
The MV alarm limits are automatically calculated until
Auto MV
Limits
is set to
Off
or until an MV limit is manually adjusted
during mechanical ventilation. The MV alarm limits are
automatically calculated based on TV and RR settings.
4.
Select
Close
.
Setting to default limits
Use
Change to Default Limits
to set alarm limits to the values set
by the Super User.
1.
Select
Alarm Setup
.
2.
Select the
Configure
tab.
3.
Select
Change to Default Limits
.
4.
Select
Close
.
Note
Aisys CS²
3-20
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...