AC.22p0
11
1
2
3
1.
Auxiliary inlet
2.
Sample gas scavenging port
3.
Outlet connection
Figure 8-2 • AGSS connections
Scavenging the ACGO sample flow
When the ACGO port is in use, a sample of the fresh gas is diverted
to the airway module or the O2 cell in the breathing system. This
sample is then measured and displays the O2 numerics on the
screen. This sample flow should be scavenged when N2O or volatile
anesthetics are delivered through the ACGO. If scavenging is not
connected, the sample flow will enter the room through the breathing
system. To connect the scavenging:
1.
Attach a circle breathing circuit to the inspiratory and expiratory
ports.
2.
Occlude the breathing circuit by connecting the wye-piece, or
patient connection, to the leak test plug located behind the
expiratory port.
3.
Check for clinically correct settings.
4.
Check the position of the Bag/Vent switch.
•
If the Bag/Vent switch is set to mechanical ventilation mode,
the bellows fills slowly with the sample flow. When the
bellows is full, the sample gas flow goes to the AGSS.
(Mechanical ventilation does not start when the ACGO
switch is set to ACGO.)
•
If the Bag/Vent switch is set to the bag mode, set the APL
valve to MIN, and attach a bag to the manual bag port. The
bag fills slowly with the sample flow. When the bag is full,
the sample flow goes to the AGSS.
8 Setup and connections
2067226-001
8-13
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...