Suction regulators (optional)
Venturi Suction Regulator
Performance Category
Pharyngeal Suction
Supply
Air or O2 from system gas supply
Drive Gas Consumption*
28 l/min with pipeline drive gas at 280 kPa
52 l/min with pipeline drive gas at 600 kPa
Maximum Vacuum*
600 mmHg with pipeline drive gas at 280 kPa
550 mmHg with pipeline drive gas at 600 kPa
Maximum Flow*
29 l/min with pipeline drive gas at 280 kPa
32 l/min with pipeline drive gas at 600 kPa
Vacuum Gauge Accuracy
+/- 5% of full scale
*Values are approximate.
Continuous Suction Regulator
Performance Category
Pharyngeal Suction
Supply
External vacuum
Maximum Vacuum*
540 mmHg with external vacuum applied of
540 mmHg and 40 l/min free flow
Maximum Flow*
39 l/min with external vacuum applied of 540
mmHg and 40 l/min free flow
Vacuum Gauge Accuracy
+/- 5% of full scale
*Values are approximate.
Aisys CS²
11-22
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...