4.
To change ventilation mode, select
Mode
. Make the change.To
change the ventilation settings, select a ventilator quick key or
More Settings
. Make the change.
5.
To change alarm settings, select
Alarm Setup
. Make the
change.
6.
To change the gas settings or the circuit type, select
Gas Setup
.
Make the change.
7.
From the
Start Case
menu, select
Start Case Now
. Gas flow
starts.
" section for information on the
Vent Mode
menu.
See the "
" section for information on the
Gas Setup
menu.
" section for information on the
Alarm Setup
menu.
End a case
Use the
End Case
menu to stop gas flow and end the patient
alarms.
1.
Set the Bag/Vent switch to Bag.
2.
Select
End Case
.
3.
Select
End Case Now
on the menu to put the system in standby
(stops the gas flow and patient alarms). The
End Case
menu
shows the gas and agent usage for the case.
Aisys CS²
3-6
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...