List of alarms
If corrective action does not resolve the alarm message, contact an
authorized service representative.
Circuit pressures and volumetric flows are measured by the
ventilator and airway gas module. If the
Data Source
is set to
Patient
the displayed waveforms and numeric information are
measured by the airway gas module. Although not displayed, the
ventilator measurements continue and if a measured value violates
an alarm setting, the appropriate alarm occurs. The value highlighted
in the parameter numeric box may not appear to have violated the
alarm setting. Changing the
Data Source
to
Vent
will display the
circuit pressures and volumetric flows measured by the ventilator.
.
Message
Priority
Cause
Action
AA, CO2
monitoring not
connected
Medium
No gas monitor is detected when the
External Gas Monitor setting in Super
User is set to No.
The monitor condition changes to
disconnected when the External Gas
Monitor setting in Super User is set to
Yes.
Connect or install a gas module in the
machine or set External Gas Monitor
setting to Yes if the system uses a
stand-alone monitor for O2, AA, and
CO2.
Agent mixture
Medium
Two different agents are detected by
the airway module and the MAC
calculation is greater than or equal to
3.
Make sure only one agent is on. Wait
approximately two minutes for the first
agent to wash out of the system.
Agent mixture
Low
Two different agents are detected by
the airway module and the MAC
calculation is less than 3.
Make sure only one agent is on. Wait
approximately two minutes for the first
agent to wash out of the system.
Agent output
inaccurate.
Contact service.
Medium
Vaporizer detected a flow meter
temperature sensor failure and is using
a default temperature value.
Contact an authorized service
representative.
Air pressure low.
Increase O2 %.
High
O2% is set to Air at 21%. Air pipeline
pressure is less than 252 kPa (36 psi)
and the air cylinder pressure dropped
below 2633 kPa (381 psi) for one
second.
Ensure the Air pipeline and cylinder are
properly connected. Increase the O2%
flow.
Air supply
pressure low
Medium
Air pipeline pressure is less than 252
kPa (36 psi) and the air cylinder
pressure dropped below 2633 kPa
(381 psi) for one second.
Ensure the air pipeline and cylinder are
properly connected.
Apnea
Medium
Apnea time delay (10-30 seconds) has
passed without a measured breath.
Apnea time delay (20-30 seconds) has
passed without a change in the
measured CO2 by at least 1%
Check for leaks in the patient circuit.
Check for patient disconnection.
Apnea >120 s
High
Apnea time exceeds 120 seconds.
Check for leaks or blockages in the
breathing circuit. Ensure the Bag/Vent
switch is in the Vent position. Check the
patient.
7 Alarms and troubleshooting
2067226-001
7-5
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...