When the liquid level indicator reaches the full mark, slowly
remove the bottle from the filling port.
Removing the bottle quickly may cause the agent to splash or
squirt out of the filling port.
--
7.
Put the cap back on the filling port.
8.
Put the cap back on the agent bottle.
Filling with Saf-T-Fill bottle
AB.60.050
AB.60.049
Figure 13-8 • Aladin2 cassette with Saf-T-Fil mechanism
Do not store filled or partially filled desflurane cassettes
above the normal working temperature of 35C/95F.
Storage at high temperature may cause the overpressure
valve to vent desflurane vapor into the ambient air.
Do not open or press the cassette filling port of the gas
connection valves with fingers or any kind of instrument.
Anesthetic agent liquid or gas may squirt into the air.
Note
WARNING
•
Aisys CS²
13-14
2067226-001