Procedures setup
Use the
Procedures
menu to set the procedures available to the
user. From this menu select the default settings for
Vital Capacity
and
Cycling
procedures.
Setting vital capacity defaults
1.
From Super User mode, select
Procedures
.
2.
Select
Vital Capacity
.
3.
Set
Show Vital Capacity
to
Yes
to show this menu item on the
user’s
Procedures
menu. Set to
No
to hide this menu item on
the user’s
Procedures
menu.
4.
Select
Pressure Hold
and set to the desired value.
5.
Select
Hold Time
and set to the desired value.
6.
Set
Show PEEP on Exit
to
Yes
to show this menu item on the
user’s
Vital Capacity
menu. Set to
No
to hide this menu item on
the user’s
Vital Capacity
menu.
7.
Select
PEEP on Exit
and set to the desired value.
8.
Select
Back
to return to the
Procedures
menu.
Setting cycling controls and settings
1.
From Super User mode, select
Procedures
.
2.
Select
Cycling
.
3.
Set
Show Cycling
to
Yes
to show this menu item on the user’s
Procedures
menu. Set to
No
to hide this menu item on the
user’s
Procedures
menu.
4.
Select the procedure to modify from the drop-down list.
5.
Select
Actions
to set up the cycling procedure. There are a
maximum of seven steps and a minimum of one step for a
cycling procedure.
Select
Adjust Settings
to make changes to the parameters for
a step.
Select
Insert Step
to add a step to the cycling procedure.
Select
Delete Step
to remove a step from the cycling procedure.
6.
Select
Close
to exit the menu.
Aisys CS²
12-16
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...