Minute volume
less than 0.1 to 99.9 liters, 0.1 liter
resolution
Breath rate
4 to 100 bpm (non-spontaneous)
2 to 60 bpm (spontaneous)
1 bpm resolution
Volume sensor type
Variable flow orifice
Oxygen
Display range
5 to 110% O2
Display resolution
1% increments
Sensor type
Galvanic fuel cell
Measurement range
0 to 100% O2
Measurement accuracy
Better than +/- 3% of full scale
Cell response time
35 seconds
Note: Response time of cell and adapters is
measured using the text method described in
ISO 7767 (1997).
Low O2 alarm range
18% to 99%
High O2 alarm setting
19% to 100% or Off
Note: Low O2 limit may not be set above high
O2 limit. High O2 limit may not be set below
the low O2 limit.
Expected cell life
Four months of shelf life (23°C room air) and
one year of normal operation.
Aisys CS²
11-38
2067226-001