Aladin cassette installation
1.
Using the liquid level indicator, check that the cassette is filled to
the appropriate level.
2.
If using Aladin2, unlock the cassette handle before installing it
into the active bay.
3.
Insert the cassette into the active bay until a click is heard. An
audible click indicates that the cassette is in position.
4.
Turn the lock on the handle to the horizontal position (Aladin2
cassettes only).
5.
The cassette is properly inserted when the agent is identified on
the display. Make sure that the displayed agent matches the
cassette.
Store the cassettes in the cassette storage bay when they are not in
use. See the "
" section for additional cassette
information.
Note
Aisys CS²
5-2
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...