Alarm ranges
The alarm names are listed in the
Primary Limits
and
More Limits
tabs on the
Alarm Setup
menu. See the "
" section
for more information on the alarm default settings.
Alarm
Range
Increment
Pmax (only high)
12-100 cmH2O
1.2 - 9.8 kPa
12 - 98 mbar, hPa
9-73 mmHg
1 cmH2O
0.1 kPa
1 mbar, 1 hPa
1 mmHg
MV High
0.5 - 30.0, Off l/min
0.5 l/min
MV Low
Off, 0.1 - 10.0 l/min
0.1 l/min
TV High
20 -1600, Off ml
20 ml
TV Low
Off, 1 - 5 ml
5 - 20 ml
20 - 1500 ml
1 ml
5 ml
20 ml
RR High
2 - 100, Off /min
1 /min
RR Low
Off, 1 -99 /min
1 /min
EtCO2 High
0.1 - 15%, Off
0.1 - 15, Off kPa
1 - 115, Off mmHg
0.1%
0.1 kPa
1 mmHg
EtCO2 Low
Off, 0.1 - 14.9%
Off, 0.1 - 14.9 kPa
Off, 1 - 114 mmHg
0.1%
0.1 kPa
1 mmHg
FiCO2 High
0.1 - 15%, Off
0.1 - 15, Off kPa
1 - 115, Off mmHg
0.1%
0.1 kPa
1 mmHg
FiO2 High
19 - 100%, Off
1%
FiO2 Low
18 - 99%
1%
EtO2 High
19 - 100%, Off
1%
EtO2 Low
Off, 1 - 99%
1%
FiIso High
0.1 - 7.0%
0.1%
FiIso Low
Off, 0.1 - 6.9%
0.1%
FiSev High
0.1 - 10.0%
0.1%
FiSev Low
Off, 0.1 - 9.9%
0.1%
FiDes High
0.1 - 20.0%
0.1%
FiDes Low
Off, 0.1 - 19.9%
0.1%
FiEnf High
0.1 - 7.0%
0.1%
FiEnf Low
Off, 0.1 - 6.9%
0.1%
FiHal High
0.1 - 7.0%
0.1%
FiHal Low
Off, 0.1 - 6.9%
0.1%
EtIso High
0.1 - 7.0%, Off
0.1%
7 Alarms and troubleshooting
2067226-001
7-13
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...