Message
Priority
Cause
Action
Backup Mode
active
Low
No spontaneous breaths in set period
of time (Backup Time) since starting
PSVPro mode.
Select a new ventilation mode. The
number of consecutive patient triggered
breaths reaches the Exit Backup
setting.
Breathing system
loose
Low
The breathing system is not latched.
Push the breathing system onto the
frame and ensure it latches.
Calibrate
(remove) flow
sensors
Low
Flow sensor calibration failure.
Calibrate the flow sensors.
Calibrate O2
sensor
Low
Calibration failure or measured O2 is
greater than 110%.
Calibrate the O2 cell. Replace O2 cell if
calibration fails.
Calibrate, dry, or
replace flow
sensors
Low
Patient volume mismatch occurred
during the last case.
Calibrate, dry or replace the flow
sensors. Start a new case.
Cannot deliver
agent setting at
set flow
Low
Cassette output flow is 6 l/min or at
maximum for more than 10 seconds.
Reduce the flow.
Cannot identify
cassette
Medium
The identification coding of the
cassette cannot be interpreted.
Try another cassette. Contact an
authorized service representative.
Cannot monitor
Air pipeline
Medium
Air pipeline pressure is invalid.
Check pipeline supply pressure.
Cannot monitor
gas supplies
Low
Hardware failure.
Contact an authorized service
representative.
Cannot monitor
O2 pipeline
Medium
O2 pipeline pressure is invalid.
Check pipeline supply pressure.
Cannot read gas
supply pressures
Medium
Transducer failure.
Contact an authorized service
representative.
Cass. overfilled.
Replace cassette
Medium
Agent level sensor indicates the
cassette is overfilled.
Try another cassette. Contact an
authorized service representative.
Check agent. Do
not fill vap in use.
Low
Cassette reporting a value of empty.
Fill the cassette.
Check cassette.
Set agent.
Medium
Cassette pressure is out of range.
Agent flow interrupted or control
failure.
Try another cassette.
Remove and reinsert the cassette.
Set the agent to the desired
concentration.
Check circuit
connections
Medium
Breaths detected in the circle circuit
while non-circuit is selected.
Check circle circuit connections and
settings.
Check D-Fend
Medium
Water trap not attached.
Check that the water trap is properly
attached to the airway module.
Check flow
sensors
Medium
System has detected an improper flow
pattern in the breathing circuit.
Ensure the flow sensors are connected
correctly.
Check sample
gas out
Medium
Possible blockage in airway module
sample gas outlet.
Check for blockage in the airway
module sample gas outlet. Remove
blockage.
Circuit leak
Medium
Vent TVexp is less than half of vent TV
insp for at least 30 seconds.
Check for leaks in the patient circuit.
Calibrate flow sensors. If problem
persists, replace flow sensors.
Aisys CS²
7-6
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...