Start a case
Use the
Start Case
menu to set the case data and to start the gas
flow.
A case can be started using default settings or using custom
settings. The default settings are configured by the Super User. See
the "
" section for information on the
Start Case
menu defaults.
Default Settings selection shows the first of five default case types
when the
Start Case
menu is accessed. Four of the default case
types are configured by the Super User. The fifth default case is Last
Case.
The
Ideal Weight
,
Age
, and
Volume Apnea Alarm
values are set to
the pre-selected settings defined by the Super User corresponding to
the case type.
Make sure that the patient breathing circuit is correctly
assembled and that the ventilator settings are clinically
appropriate before starting ventilation. Incorrect breathing
circuit assembly and incorrect ventilator settings can
injure the patient.
Make sure that the preset alarm limits are appropriate for
the patient before starting ventilation. Incorrect alarm
settings can injure the patient.
Volume Apnea Alarm
is not shown on the
Start Case
menu when
the
Volume Apnea Selection
is set to
Disable
in the Super User
settings.
The
TV for Ideal Body Weight
menu item from the
Patient
Demographics
menu can only be accessed when the ventilation
mode is set to
VCV
,
PCV-VG
,
SIMV VCV
, and
SIMV PCV-VG
. Use
this setting for breath rate and tidal volume calculations based on the
set patient weight.
Minimum Alveolar Concentration
The adjusted Minimum Alveolar Concentration (MAC) is calculated
based on the patient age entered in the
Start Case
menu or the
Patient Demographics
menu. The default patient age of the
selected case type is used if no patient age value is entered.
The MAC value is calculated from the exhaled gas concentration and
the related affects based on the age of the patient. Typically,
younger patients have better liver function and can clear a drug
faster, resulting in a higher MAC value. The MAC calculation used is
based on the Eger formula. When two agents are detected, the MAC
values of each agent are added together. The MAC value range is
0.0 to 9.9.
WARNING
•
Note
Note
Aisys CS²
3-4
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...