Passive scavenging
Negative pressure relief
0.3 cmH2O
Outlet connector
30 mm male taper ISO
Active scavenging
Disposal system
type
Outlet
connector*
Flow range
Pressure
Adjustable
DISS EVAC
Up to 30 l/min
not applicable
High flow, low
vacuum
BSI 30 mm
threaded
50 to 80 l/min
not applicable
Low flow, high
vacuum
DISS EVAC
36 +/- 4 l/min
305 mmHg (12
inHg) minimum
vacuum
Low flow, low
vacuum
12.7 mm hose
barb
36 +/- 4 l/min
not applicable
Low flow, low
vacuum
25 mm hose barb 40 to 50 l/min
not applicable
Low flow, low
vacuum
30 mm ISO taper 40 to 50 l/min
not applicable
*Other market-specific connectors may be available.
Particle filter at the outlet has a pore size of 225 microns. All flow data uses
a new filter.
11 Specifications and theory of operation
2067226-001
11-15
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...