Removing the bottle quickly may cause the agent to splash or
squirt out of the filling port.
--
7.
Put the cap back on the filling port.
8.
Remove the bottle adapter from the anesthetic agent bottle. Put
the cap back on the agent bottle.
Filling with Quik-Fil system
AB.60.041
Figure 13-7 • Aladin2 cassette with Quik-Fil system
Do not open or press the cassette filling port or the gas
connection valves with fingers or any kind of instrument.
Anesthetic agent liquid or gas may squirt into the air.
1.
Remove the cassette from the machine, and place it on a
horizontal surface. Make sure that the valve pins on the back of
the cassette do not come in contact with the machine or any
other object.
2.
Remove the yellow protective cap from the anesthetic agent
bottle. Check that the filling mechanism on the bottle is not
damaged.
3.
Remove the filling port cap from the cassette by turning the cap
counterclockwise.
4.
Align the bottle nozzle keys with the index slots in the filling port.
5.
Push the agent bottle firmly into the filling port. Let the agent
flow into the cassette.
6.
Watch the liquid level indicator.
Note
WARNING
13 Vaporizer cassettes
2067226-001
13-13
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...